腹腔鏡手術研究向けオープンソースda Vinci X/Xi器具モジュール
An Open-Source da Vinci X/Xi Instrument Module for Laparoscopic Surgery Research
da Vinci X/Xiの手術器具をUR5/UR5eロボットアームで駆動し、電気手術の通電制御と記録を可能にするオープンソースモジュールを開発し、ベンチ・摘出組織で検証した。
著者: Douglas McDonough, Justin Liao, Jiawei Ge, Antony Goldenberg, Yvonne Zhang, Mackenzie A Eagleson, Ravi Prakash, Axel Krieger, Chunwoo Kim
分類: cs.RO
原文アブストラクト
Studying electrosurgical tasks on a research robot requires control of both instrument motion and energy activation, along with a record of when each action occurs. This paper presents an open-source module for da Vinci X/Xi instruments on a UR5/UR5e robotic arm. Five motors transmit torque through timing belts to spring-loaded drive discs. An electrosurgical interposer accepts software commands, gives the surgeon's foot pedals priority, and records the source of each activation request. In bench testing, the interposer recorded all 12 activation and deactivation events from six pedal presses. Ex vivo demonstrations included monopolar and bipolar coagulation of splenic tissue and monopolar cutting during gallbladder dissection. Optical tracking of the Fenestrated Bipolar Forceps showed that calibration brought pitch and yaw gains from 0.904 and 1.14 to 1.00 and 1.02, respectively. The module enables researchers to record instrument motion, video, and the timing and source of electrosurgical commands during laparoscopic tasks.