滅菌ドレープ外のUWBアンテナとIMU、トロカール位置拘束を組み合わせ、光学情報なしで手持ち腹腔鏡器具の先端位置を推定する手法を提案。

著者: Jinseok Lee, Dongho Yee, Minsung Kim, Younghoon Noh, Seonho Shim, Juahn Oh, Yechan Seo, Jiyul Lee, Seong Jeong, Hyuk Choi, Hyoun-Joong Kong

分類: cs.RO

原文アブストラクト

A surgical robot policy needs to know where each instrument's working end sits relative to the camera and to the other instrument, yet a laparoscopic operation leaves only the endoscope video, and a draped instrument hides every optical path on itself. We estimate the tool tips from distances and an IMU alone. The distances are measured by ultra-wideband carrier phase between antenna nodes on the handle side of the instruments: one per hand-held instrument, one on the endoscope, all outside the sterile barrier. Take the endoscope antenna as the reference. The two instrument antennas carry six coordinates while the three pairs give three distances, so the problem is short by three, and averaging cannot close that gap because what is missing is information, not precision. Carrier phase adds one more unknown per pair, an integer that leaves distance fixed only modulo lambda/2 = 23.1 mm. The trocar settles both difficulties: each shaft passes through a port whose position is known and whose direction the on-board IMU measures, so its antenna keeps a single degree of freedom, the insertion depth. Two unknowns now face three distances -- two fix the depths and the third is left over, and that spare equation exposes a wrong integer or a wrong mounting offset instead of absorbing it. The same solve returns both tool tips in the endoscope frame, without the second carrier frequency that integer resolution usually needs, so the link never leaves a single PLL lock. In an 11-hole phantom among metal instruments the three distances hold sigma = 0.10-0.33 mm at 35-46 Hz, a sterile drape pressed onto the antennas leaves the link unchanged where an optical path would be blocked, and the logger was carried into a live rabbit appendectomy.

PR本紙発行元 EmplifAI