Comparing Commercial Depth Sensor Accuracy for Medical Applications

Comparing Commercial Depth Sensor Accuracy for Medical Applications

シェア:XThreadsFacebookLINEはてブBluesky

著者: Pit Henrich, Maximilian Weiherer, Franziska Hansen, Bernhard Egger, Franziska Mathis-Ullrich

分類: cs.RO, cs.CV

原文アブストラクト

Depth estimation has numerous medical and surgical applications. We benchmark four depth sensors on a porcine bone specimen, a porcine belly specimen, and a silicone kidney phantom using stylus-sampled references. These objects contain several real-world challenges, including homogeneous surfaces, specular surfaces, and subsurface scattering. The comparison includes stereo, structured-light, and time-of-flight sensors at a distance of approximately 50 cm. Specifically, the Intel RealSense D405 (Intel RealSense, United States), PMD Flexx2 (pmdtechnologies, Germany), Stereolabs ZED 2i (Stereolabs, France), and Zivid 2M+ 60 (Zivid, Norway) are compared. The Zivid 2M+ 60 performed best across all objects and metrics considered in this work. The ZED ranked second for real tissue, but last on the phantom.